Horizontal Rectangular Steam Sterilizer (HRA55)
Specification & Features
SIZE: 600mm X 900mm X 1500mm.
CONTROL:
- Multi-port operating valve, control for the entire cycle of the operation.
- Permits two exhaust speeds, fast and slow.
- Easy to read compound gauge for the chamber and pressure gauge for the jacket.
- Smooth operation.
SAFETY FEATURES:
- Self-locking safety door cannot be opened while chamber is under pressure.
- Gauge glass tube with automatic closing safety valve.
- Heat resistant silicone door gasket.
- Pressure switches economies power consumption.
- Low water protection system, to protect the electric heaters.
- Temperature gauge indicators.
- Vacuum breaker.
- Safety valve for the jacket/ boiler.
- Hydraulically test to 2.5 times the designed working pressure.
- Powerful ejector for drying.
- Insulated with resin bond fibre glass wool covered with SS 304 sheet.
CONSTRUCTION:
- Chamber & Jacket – The material use for chamber is of heavy duty 304/316 L grade stainless steel & jacket SS 304.
- Doors: The doors are constructed of heavy duty 304/316 grade SS, both single & double door models are available.
- Piping:– The pipes & fittings are of stainless steel and brass. The valve is ball type.
- Mounting:- The sterilizer can be supplied in one of following mounting alternatives, built in wall mounted, A flush mounted panel to isolate the sterilize side from the non-sterile side, free standing with & without cabinet.
- Panel:- All sterilizer is fitted with powder coated/ stainless steel panels, special panel design can be supplied according to customer requirements(Optional).
DOORS:
- Depending on the process and space availability, we provide you with the following options in selecting the door the autoclave single / double door manually operated hinged door.
DOOR SAFETY FEATURES:
- A special safety device feature prevents the operator from opening all door types in the flowing circumstances.
- When the chamber is pressurized.
- If the temperature in the chamber is above safety level.
HEATING:
- The sterilizer is heated by steam. The steam can be supplied from the building is central source, an electric stainless steel steam generator 6KW X 6 Nos Heating elements.
- Power: 36Kw, Three Phase
- Loading (Optional): Pull-Out trays/ carriages and transfer trolley for loading and unloading process.